8.9 Message Reference
178
8.9.2 Unique Commands
Setting and Query of the number of measurements to
Average
Syntax
Command
Query
Response
:AVERaging
<Numeric Value>
:AVERaging?
<Numeric Value> = 1 to 256
(NR1)
Explanation
Command Set the number of measurements to average for the average mea-
surement value.
When the number of average measurements is set, the average func-
tion will not be automatically set to ON.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric value can be handled.
Query
Returns the setting of the number of measurements to average as
NR1 numeric value.
Example
Command
:AVERaging 32
Set the number of measurements to average to 32.
Query
Response
:AVERaging?
:AVERaging 32
(when HEADER ON)
32
(when HEADER OFF)
The number of measurements to average is set at 32.
Note
The command receives 1, but the response returns OFF.
Setting and Query of Average
Syntax
Command
Query
Response
:AVERaging:STATe
<ON/ OFF>
:AVERaging:STATe?
<ON/ OFF>
Explanation
Command Sets the ON/ OFF setting of average function.
Query
Returns ON or OFF for the setting of average function.
Example
Command
:AVERaging:STATe ON
Enables the average function.
Query
Response
:AVERaging:STATe?
:AVERaging:STATE ON
(when HEADER ON)
ON
(when HEADER OFF)
The average function is enabled.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......