8.9 Message Reference
199
8
Setting and Query of Open Compensation Values
Syntax
Command
Query
Response
:CORRection:OPEN:DATA
<Compensation values 1>
,
<Com-
pensation values 2>
:CORRection:OPEN:DATA?
<Compensation values 1>
,
<Compensation values 2>
<Compensation values 1>
= -99.9999E9 to 99.9999E9 (NR3)
<Compensation values 2>
= -99.9999E9 to 99.9999E9 (NR3)
Explanation
Command Sets the open compensation value for current measurement condi-
tions (frequency, level).
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns as a numeric value the open compensation value for current
measurement conditions (frequency, level).
Note
• The forwarding format for making settings and reading data with
this command follows the :CORRection:OPEN:DATA:FORMat set-
ting.
When the forwarding format is ZPH
<Compensation values 1> : Z
( Z
0 an execution error)
<Compensation values 2> : PH
( PH
-180, PH
180 an execution error)
When the forwarding format is GB
<Compensation values 1> : G
<Compensation values 2> : B
When the forwarding format is CPG
<Compensation values 1> : Cp
<Compensation values 2> : G
• When the compensation value is-1E-21 to 1E-21 the setting value
will be 0.
Example
Command
:CORRection:OPEN:DATA:FORMat GB
:CORRection:OPEN:DATA -1.56789E-11,8.91234E-11
When the forwarding format setting is GB, the open compensation
value for the current measurement conditions will be set to
-1.56789E-11,8.91234E-11
(G,B).
Query
:CORRection:OPEN:DATA?
:CORRECTION:OPEN:DATA -1.56789E-11,8.91234E-11
(when HEADER ON)
When the forwarding format setting is GB, the open compensation
value for the current measurement conditions is set to
-1.56789E
-11,8.91234E-11
(G,B).
Note
• If the value forwarded to this device when converted to G and B is
outside of the -99.9999E9 to 99.9999E9 range an execution error
will occur.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......