5.1 Comparator Function
67
9
5
Measurement
method 2
Make settings using a reference value and the upper limit
and lower limit values (
% setting)
(To make settings using a reference value and the upper limit and lower limit val-
ues refer to "Measurement method 1" (p. 64))
1.
Press
to light up COMP and enter comparator measurement
mode.
2.
Press .
3.
Press
to enter "C" reference value setting mode.
4.
Enter the C reference value using the arrow keys or the numeric
keypad.
(If you enter a number, each digit moves one place to the right.)
Settable Range : -199999 to 999999 (The
%
setting cannot be set to "0")
Move to the digit.........
and
Change the number ...
and
Setting a value lower than the lower limit value will not produce an error, but
accurate judgment will not be possible.
Set a count value for the reference value.
At time of shipment, it is set to ” 100000 ”.
(
Light up
)
To make settings from an optional setting
mode, press
while ”C” or ”d” is
flashing to select the setting mode, then
press .
Setting modes: C reference value
C
upper limit and lower limit values
D ref-
erence value
D upper limit and lower
limit values
C reference value...
(SUB display area)
Light up
Flash
(SUB display area)
Light up
Flash
Reference value
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......