5.2 BIN Measurement Function (Model 3504-50, 3504-60 only)
82
Measurement
method 2
Make settings using the reference value upper limit and
lower limit values (
% setting)
(To make settings using a reference value and the upper limit and lower limit val-
ues refer to "Measurement method 1" (p. 78))
1.
Press
to light up BIN and enter the BIN measurement mode.
2.
Press .
3.
Press
to enter the C reference value setting mode.
4.
Enter the C reference value using the arrow keys or the numeric
keypad.
(If you enter a number, each digit moves one place to the right.)
Settable Range: -199999 to 999999 (Cannot be set to 0.)
Move to the digit ........
and
Change the number ..
and
The reference value is set with the count value. At time of shipment it is set to
”100000”.
(
Light up
)
To make settings from an optional setting
mode, press
while "C" is flashing to
select the setting mode, then press
.
Setting modes: C reference value
BIN1
to 14 upper limit and lower limit values
D reference value
D upper limit and
lower limit values
C reference value....
(SUB display area)
Light up
Flash
(SUB display area)
Flash
Light up
Reference value
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......