5.1 Comparator Function
73
9
5
Comparator Measurement Results ________________________________
• The measurement conditions of comparator measurement mode use the mea-
surement conditions of normal measurement. Set the comparator mode mea-
surement conditions in normal measurement mode.
See
3.3 "Setting the Measurement Conditions" (p. 26)
• The upper limit and lower limit value numbers are not checked for which is
greater and smaller. Therefore please take care when setting them, as correct
judgments cannot be made if they are set in the opposite order.
Judgement results
display
Each of the judgment results for C and D is displayed in the comparator judg-
ment result display area.
The comparator judgment is not performed for parameters with the upper limit
and lower limit value set to OFF.
Measurement value judgment
Upper limit value
Lower limit value
HI
IN
LO
Upper limit and
lower limit value
settings
When the measurement value is larger than the
upper limit value
When the measurement value is within the
range of the upper limit and lower limit values
When the measurement value is smaller than
the lower limit value
When the measurement value is above the
measurement range (OVER FLOW)
When the measurement value is below the
measurement range (UNDER FLOW)
Upper limit
Lower limit
HI
LO
Measurement
range
Measurement
range
Measurement value judgment
Judgment
order
Judgement results
Meaning
Solution
1
The HI LED lights up. Measurement error
See
The HI LED lights up. The measurement value is above the high limit
value range in the current range.
Switch to an appropriate
range.
See
The LO LED lights up. The measurement value is below the low limit
value range in the current range.
2
The LO LED lights up. The measurement value is lower than the lower
limit value.
3
The HI LED lights up. The measurement value is higher than the up-
per limit value.
4
The IN LED lights up. The measurement value is within the setting
range.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......