1.4 Names and Functions of Parts
14
MAIN display area ERROR display
When the
3504-40, 3504-50 and 3504-60 C HiTesters detects a measurement abnormality, an error message is dis-
played in the MAIN display area.When a measurement error occurs, the device's condition is displayed in order of priority
rank in the MAIN display area.When an error display occurs, the comparator and BIN measurement judgment results will
be HI and OUTOF-B
INS.
Priority
rank
MAIN display
Error content
EXT I/O
Solution
-
OPEN compensation error
Displayed when the OPEN
compensation value is less than
1 k
.
See
-
• Put the measurement terminals in an open
state. (Short circuit the H
CUR
terminal to the
H
POT
terminal, and the L
CUR
terminal to the
L
POT
terminal.)
• Use the shielding process as a countermeasure
against external noise.
• Earth the device.
• Check to see if the measurement cable is broken.
SHORT compensation error
Displayed when the SHORT
compensation value is more
than 1 k
See
-
• Short the measurement terminals.
• Check to see if the measurement cable is broken.
LOAD compensation error
Displayed when the LOAD com-
pensation value is outside of
range.
See
-
• Perform compensation again after setting to the
appropriate range.
-
Output current abnormality
Displayed when low imped-
ance elements were con-
nected for more than 10 min.
in range
7 or 8.
-
• Put the measurement terminals in an open state
and then press
.
• In Range 7 and 8, do not leave low impedance
elements (1
or less) connected. It may result
in damage on the main unit.
-
Normal
measurement value
ERR LED light on
Acquiring self calibration
abnormal value
See
ERR
output
Normal
evalua-
tion
• The device is acquiring an abnormal self cali-
bration value. Reconnect the instrument to be
tested correctly and carry out self-calibration
again.
C
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......