7.4 About Measurement Times
129
9
7
End of Measure-
ment Signal
EOM
The output time (
T4
) for an end of measurement signal (EOM) can be obtained
by the following equation.
T4 = A(1 or 2) + B + C + D
A
Measurement time during normal measurement mode, no open/short circuit
compensation, HOLD range (contact check: OFF, trigger simultaneous output
function: disabled)
(1) No request for self calibration
(Allowable tolerance:
±
5%
±
0.5 ms)
(2)
Request for self calibration
B
The calculation time differs depending on whether or not there is open circuit,
short circuit, or load compensation
C
The calculation time differs if comparator is executed.
D
The calculation time differs if BIN is executed. (Only for 3504-50, 3504-60)
Measurement Frequency
Measurement Speed
FAST (ms)
NORM (ms)
SLOW (ms)
120 Hz
10.0
37.5
146.0
1 kHz
2.0
5.5
29.5
Measurement Frequency
Measurement Speed
FAST (ms)
NORM (ms)
SLOW (ms)
120 Hz
25.8
43.2
148.2
1 kHz
4.8
7
30
Open Circuit, Short Circuit and
load Circuit Compensation
(ms)
No
0.0
Yes
Max. 0.4 Each
Measurement Mode
(ms)
Normal measurement mode
0.0
Comparator Measurement Mode
MAX 0.4
Measurement Mode
(ms)
Normal measurement mode
0.0
BIN Measurement Mode
MAX 0.4
The output time for an analog measurement signal (INDEX) will be delayed by
the execution of the contact check. Accordingly the output time for an end of
measurement signal (EOM) will be delayed by the increase of delay time of
INDEX output.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......