8.13 Device Document Requirements (Only for Models 3504-50, 3504-60)
252
Standard implementation method related information based on the IEEE 488.2
standard
(1) Functionality of IEEE 488.1 interface function
This is included in 8.2.2 "GP-IB Specifications (Only for Models 3504-50, 3504-
60)" (p. 133).
(2) Explanation of operation when the address is set to other than a value
from 0 to 30.
Such a setting is not possible.
(3) Recognition of change to the address initially set by the user
The change to the address is recognized when the address is changed.
(4) Explanation of the device settings at power on
The status information is cleared. Other information is backed up. However, the
header and response message terminator are initialized.
(5) Description of message exchange options
• Capacity and operation of input buffer
This is included in8.6.2 "About the Output Queue and Input Buffer" (p. 146).
• Queries that return multiple response message units
:BIN:FLIMit:COUNt? ............................................3
:BIN:FLIMit:DEViation? .......................................3
:BIN:SLIMit:COUNt? ...........................................3
:BIN:SLIMit:DEViation? .......................................3
:COMParator:FLIMit:COUNt? .............................2
:COMParator:FLIMit:DEViation? .........................3
:COMParator:SLIMit:COUNt? .............................2
:COMParator:SLIMit:DEViation? .........................3
:CORRection:DATA?............................................4
:CORRection:LOAD:DATA? ................................2
:CORRection:LOAD:REFerence? .......................2
:CORRection:OFFSet:DATA? ..............................2
:CORRection:OPEN:DATA?.................................2
:CORRection:SHORt:DATA? ...............................2
:MEASure?.................................................. 2, 3, 5
:MEMory? ...........................................1 to 32000
:SSOurce:WAIT? .................................................2
• Queries that result in the creation of a response once the syntax is checked
All queries result in the creation of responses once the syntax is checked.
• Use/non use of queries that result in a response being created upon being
read
Queries that result in a response being created when they are read by the
controller are not used.
• Use/non use of a coupling command
There is no such command.
8.13 Device Document Requirements
(Only for Models 3504-50, 3504-60)
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......