3.2 Measurement Example
24
The following example shows a measurement operation using the 3504-40, 3504-50, 3504-60
.
3.2 Measurement Example
Example
The
9263 SMD Test Fixture
is used for the measurement
of multilayer ceramic capacitors.
Necessary tools
• 3504-40, 3504-50, 3504-60
• 9263 SMD Test Fixture
• Sample to be measured: Multilayer ceramic capacitor
Measurement
Conditions
See .
2
Connect the 9263 SMD Test Fixture (Option).
1
Connect the 9263 SMD Test Fixture to the measure-
ment terminal.
For the connection method, refer to the instruction man-
ual supplied with the fixture.
Set the measurement conditions.
2
Using the keys on the operating panel, set the mea-
surement conditions as shown at left.
Make other settings as necessary.
See
4.1 "Open Circuit Compensation and Short Circuit
See
4.2 "Load Compensation" (p. 47)
See
4.3 "Offset Compensation" (p. 54)
See
4.4 "Self Calibration" (p. 58)
See
6.1 "Setting the Average Function" (p. 91)
See
6.2 "Trigger Delay Setting" (p. 93)
See
3.3.7 "Trigger Signal" (p. 36)
The open circuit compensation, short circuit compen-
sation, load circuit compensation and self calibration
settings improve measurement accuracy.
MODE
Measurement mode........ NORM
FREQ
Frequency......................... 1 kHz
LEVEL
Measurement signal level ..... 1 V
SPEED
Measurement speed ....... NORM
CIRCUIT
Equivalent-circuit mode .. AUTO
RANGE
Measurement range......... AUTO
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......