4.3 Offset Compensation
54
Compensates for discrepancy from the correct value by subtracting an
optional input value from measurement results. This function can compen-
sate for the discrepancy found when measuring a fixed sample, or enable
interchangeability of measuring instruments when measuring a single sam-
ple.
Performing offset compensation _________________________________
1.
In normal measurement mode, press
.
The upper part of the MAIN display area displays the menu contents and the
lower part displays the setting information.
(Refer to "Menu display organization" (p. 13) for menu order)
This setting cannot be changed in comparator mode and BIN mode.
2.
Use
or
to select the “oFFSEt” menu item.
The state becomes as follows.
4.3 Offset Compensation
(MAIN display area)
(Offset compensation setting screen)
Flash
(SUB display area)
Light up
Summary of Contents for 3504-40 C HiTester