4.3 Offset Compensation
57
4
9.
Input the D offset compensation value with the arrow keys or the
numeric keypad.
(If you enter a number, each digit moves one place to the right.)
Settable Range of D : -1.99999 to 1.99999
Move to the digit.........
and
Change the number ...
and
10.
Press
to confirm the D offset compensation value.
The following will be displayed on the upper portion of the MAIN display when
“oFF” is selected.
• 3504-40, 3504-50......“Low C“ (Low C reject function setting screen)
• 3504-60.....................“Cont“ (Contact check function setting screen)
11.
Press .
The unit returns to normal measurement mode.
• At time of shipment, the C and D offset compensation values are set to "0".
• The C and D offset compensation values are both divided into separate data
depending on the measurement frequency (120 Hz, 1 kHz).Set the offset
compensation value for the presently set measurement frequency.