8.9 Message Reference
192
Setting and Query of Upper Limit and Lower Limit Values of First Parame-
ter for Comparator Function in Count Value Mode
Syntax
Command
Query
Response
:COMParator:FLIMit:COUNt
<Lower Limit Value>,<Upper Limit
Value>
:COMParator:FLIMit:COUNt?
<Lower Limit Value>,<Upper Limit Value>
<Lower Limit Value>
= OFF/ Numeric Value from -199999 to 999999 (NR1)
<Upper Limit Value>
= OFF/ Numeric Value from -199999 to 999999 (NR1)
Explanation
Command Sets the upper limit and lower limit values of the first parameter for the
comparator function in count value mode.
A numeric value in NRf format is accepted but non significant digits
are rounded off so the numeric.
Query
Returns data for the upper limit and lower limit value settings of the
first parameter for the comparator function in order of lower limit value
and upper limit value.
Example
Command
:COMParator:FLIMit:COUNt 112345,123456
Sets 112345 for the lower limit value and 123456 for the upper limit
value of the first parameter in count value mode.
Query
Response
:COMParator:FLIMit:COUNt?
:COMPARATOR:FLIMIT:COUNT 112345, 123456
(when HEADER ON)
112345,123456
(when HEADER OFF)
112345 is set for the lower limit value and 123456 is set for the upper
limit value of the first parameter in count value mode.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......