4.2 Load Compensation
53
4
Setting Load Circuit Compensation ON/ OFF _______________________
1.
In normal measurement mode, press
.
The state becomes as follows.
2.
Press
to turn the load circuit compensation ON or OFF.
Pressing
or
causes the display to change in the following manner.
3.
Press
to confirm the load circuit compensation ON or
OFF
.
The state of the load compensation LED becomes as follows.
(Flash)
(MAIN display area)
Flash
“AdJuSt”
Delete the disabled load circuit compensation values,
acquire the load circuit compensation value at the cur-
rently set measurement frequency and enable the load
circuit compensation value.
“SEt”
Set the standard value at the currently set measure-
ment frequency.
“oFF”
Disable load circuit compensation value at the enabled
frequency.
“on”
Enable all load circuit compensation value at all dis-
abled frequencies.
(Light up)
(Off)
When select the “oFF”
(Off)
(Light up)
When select the “on”
• Load circuit compensation coefficient will not be enabled even if “on” is
selected before the acquisition or setting of the load circuit compensation
coefficient.
• Press
if not turning load compensation ON or OFF.
The unit returns to normal measurement mode.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......