8.9 Message Reference
190
Setting and Query of Contact Check Function
Syntax
Command
Query
Response
:CONTact:VERify
<Character>
:CONTact:VERify?
<Character> = OFF/ BEFore/ AFTer/ BOTH
OFF
: Contact check will not be carried out.
FRONt : Carry out contact check before measurement.
BACK : Carry out contact check after measurement.
BOTH : Carry out contact check before measurement and after mea-
surement.
Explanation
Command Set the contact check function.
Query
Return setting for contact check function using words.
Example
Command
:CONTact:VERify AFTer
Set to carry out the contact check after measurement.
Query
Response
:CONTact:VERify?
:CONTACT:VERIFY AFTER
(when HEADER ON)
AFTER
(when HEADER OFF)
Set to carry out contact check after measurement.
Note
When the contact check function is set to “bEForE” or “both, the trig-
ger simultaneous output function will automatically turn to “ON”.
Setting and Query of ON/ OFF Setting of Comparator Function
Syntax
Command
Query
Response
:COMParator
<ON/ OFF>
:COMParator ?
<ON/ OFF>
Explanation
Command Sets the ON/ OFF setting of the comparator function.
Query
Returns ON or OFF for the setting of the comparator function.
Example
Command
:COMParator ON
Sets the comparator function to ON.
Query
Response
:COMParator?
:COMPARATOR ON
(when HEADER ON)
ON
(when HEADER OFF)
The comparator function is set to ON.
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......