2.2 Checking the Power Voltage
18
The power voltage specification of the unit is set as specified when the unit was
ordered.
You can select from 100 V, 120 V, 220 V, and 240 V.
You can determine which voltage is set by checking the positions of the voltage
selectors.
Refer to the diagram between the voltage selectors.
2.2 Checking the Power Voltage
• Before turning the unit on, make sure the supply voltage matches that
indicated on the its power connector. Connection to an improper sup-
ply voltage may damage the unit and present an electrical hazard.
• The power of the unit can be changed with the voltage selectors. To
avoid an electric accident, use the unit with the voltage selectors set to
a voltage value that matches the voltage to be used.
• Make sure the power is off when you change the voltage with the volt-
age selectors. Changing the power voltage when the power is on may
result in damage to the unit or an electric accident.
• The maximum rated power is 110 VA.
• Replace the fuse only with one of the specified characteristics and volt-
age and current ratings. Using a non-specified fuse or shorting the
fuse holder may cause a life-threatening hazard.
Fuse type: 100 V 120 V setting: 250 V F1.0AL
20 mm x 5 mm dia
220 V 240 V setting: 250 V F0.5AL
20 mm x 5 mm dia
See
10.2 "Replacing the Power Fuse" (p. 265)
Voltage
Position of Left Voltage
Selector
Position of Right Voltage
Selector
100 V
(Right side)
(Right side)
120 V
(Right side)
(Left side)
220 V
(Left side)
(Right side)
240 V
(Left side)
(Left side)
Back
In the diagram, the voltage
value is 100 V because
both the left and right volt-
age selectors are set to the
right side.
Voltage selectors
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......