10.2 Replacing the Power Fuse
265
10
10.2 Replacing the Power Fuse
• To avoid an electric accident, be sure to turn the power switch off and
disconnect the power cord before replacing the fuse or switching the
power voltage.
Afterwards, be sure to check that the power voltage set for the power
switch with power selectors on the rear of the unit and the voltage to be
used match before reconnecting the power cord.
• Replace the fuse only with one of the specified characteristics and volt-
age and current ratings. Using a non-specified fuse or shorting the
fuse holder may cause a life-threatening hazard.
Fuse type: 100 V 120 V setting: 250 V F1.0AL
5 mm x 20 mm
220 V 240 V setting: 250 V F0.5AL
5 mm x 20 mm
• Prior to shipment, the power voltage is set and the fuse designated for
that power voltage is included (same applies for the spare fuse). If you
want to set another power voltage, be sure to replace the fuse with one
of the designated rating.
• If you want to use a power voltage other than one indicated, set the corre-
sponding power voltage setting shown below.
Set the power voltage to 110 V
120 V/ 200V
220 V/230 V
240 V
See
2.2 "Checking the Power Voltage" (p. 18)
Screwdriver
Fuse holder
Fuse:
5 mm x 20 mm
Replacing the Power Fuse
Rear panel of the unit
Tools to Prepare: Flat blade screw-
driver
1.
Turn off the power switch and dis-
connect the power cord.
2.
Align the flat blade screwdriver
with the fuse holder securing part
of the power inlet and then remove
the fuse holder by pushing the
handle of the screwdriver toward
the opposite side of the unit.
3.
Replace the power fuse with a
fuse of the designated rating.
4.
Reinsert the fuse holder in the
power inlet.
Removing the Fuse Holder
Summary of Contents for 3504-40 C HiTester
Page 2: ......
Page 22: ...1 4 Names and Functions of Parts 16 ...
Page 28: ...2 5 Turning the Power On and Off 22 ...
Page 40: ...3 3 Setting the Measurement Conditions 34 OF UF Evaluation Flow Chart ...
Page 66: ...4 4 Self Calibration 60 ...
Page 96: ...5 2 BIN Measurement Function Model 3504 50 3504 60 only 90 ...
Page 136: ...7 4 About Measurement Times 130 ...
Page 286: ...Appendix 7 Initial Settings Table A14 ...
Page 289: ......
Page 290: ......
Page 291: ......
Page 292: ......