SC600 Oscilloscope Calibration Option
Verification Tables
9
9-37
Table 9-2. SC600 Option DC Voltage Verification (cont.)
(1 M
Ω
output impedance unless noted)
Nominal Value (V dc)
Measured Value (V dc)
Deviation (V dc)
1-Year Spec. (V dc)
-0.499
0.0002895
0.5
0.00029
-0.5
0.00029
1.35
0.000715
-1.35
0.000715
2.19
0.001135
-2.19
0.001135
2.2
0.00114
-2.2
0.00114
6.6
0.00334
-6.6
0.00334
10.99
0.005535
-10.99
0.005535
11 0.00554
-11
0.00554
70.5
0.03529
-70.5
0.03529
130
0.06504
-130
0.06504
6.599 (50
Ω
)
0.0165375
Table 9-3. SC600 Option AC Voltage Amplitude Verification
(1 M
Ω
output impedance unless noted)
Nominal Value
(V p-p)
Frequency (Hz)
Measured Value
(V p-p)
Deviation (V p-p)
1-year Spec.
(V p-p)
0.001 1000
0.000041
-0.001 1000
0.000041
0.01 1000
0.00005
-0.01 1000
0.00005
0.025 1000
0.000065
-0.025 1000
0.000065
0.11 1000
0.00015
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...