Maintenance
Performance Tests
7
7-21
Table 7-15. Verification Tests for Frequency
Range, Normal
Output, V
Output, Normal,
V
Frequency Lower
Limit
[1]
Upper
Limit
[1]
3.29999 3.00000
119.00
Hz
118.99602 Hz
119.00398 Hz
120.0 Hz
119.99600 Hz
120.00400 Hz
1000.0 Hz
999.974000 Hz
1000.026000 Hz
100.00 kHz
99.99750000 Hz
100.00250000 Hz
[1]
Frequency accuracy is specified for 1 year.