SC600 Oscilloscope Calibration Option
Verification Tables
9
9-45
Table 9-10. SC600 Option Leveled Sine Wave Verification: Flatness (cont.)
Nominal Value
(V p-p)
Frequency
Measured Value
(V p-p)
Deviation (V p-p)
1-Year Spec.
(V p-p)
0.01 50
kHz
na na
0.01 30
MHz
0.00025
0.01 70
MHz
0.00025
0.01 120
MHz
0.0003
0.01 290
MHz
0.0003
0.01 360
MHz
0.0005
0.01 390
MHz
0.0005
0.01 400
MHz
0.0005
0.01 480
MHz
0.0005
0.01 570
MHz
0.0005
0.01 580
MHz
0.0005
0.01 590
MHz
0.0005
0.01 600
MHz
0.0005
0.025 50
kHz
na
na
0.025 30
MHz
0.000475
0.025 70
MHz
0.000475
0.025 120
MHz
0.0006
0.025 290
MHz
0.0006
0.025 360
MHz
0.0011
0.025 390
MHz
0.0011
0.025 400
MHz
0.0011
0.025 480
MHz
0.0011
0.025 570
MHz
0.0011
0.025 580
MHz
0.0011
0.025 590
MHz
0.0011
0.025 600
MHz
0.0011
0.039 50
kHz
na
na
0.039 30
MHz
0.000685
0.039 70
MHz
0.000685
0.039 120
MHz
0.00088
0.039 290
MHz
0.00088
0.039 360
MHz
0.00166
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...