SC600 Oscilloscope Calibration Option
Remote Commands and Queries
9
9-33
Parameters: (None)
RANGE
Programs the instrument range in PULSE, MEAS Z, and OVERLD modes.
Parameters:
TP0DB
Sets the range to 2.5 V in pulse mode.
TP8DB
Sets the range to 1.0 V in pulse mode.
TP20DB
Sets the range to 250 mV in pulse mode.
TP28DB
Sets the range to 100 mV in pulse mode.
TZ50OHM
Sets the impedance to 50
Ω
in Meas Z mode.
TZ1MOHM
Sets the impedance to 1 M
Ω
in Meas Z mode.
TZCAP
Sets the impedance to cap in Meas Z mode.
TOLDC
Sets the instrument to DC in Overload mode.
TOLAC
Sets the impedance to AC in Overload mode.
Example:
RANGE TP20DB
Edge Function Commands
TDPULSE
(IEEE-488, RS-232, Sequential)
Turns tunnel diode pulse drive on/off in EDGE mode.
Parameters:
ON
(or non-zero) or
OFF
(or zero)
Example:
TDPULSE ON
Shows the tunnel diode pulse drive setting in EDGE mode.
Parameters: None
Response:
1 if ON, 0 if OFF.
Marker Function Commands
TMWAVE
(IEEE-488, RS-232, Sequential)
Selects the waveform for MARKER mode.
Parameters:
SINE
Sine wave (2 ns to 15 ns)
SPIKE
Triangular/sawtooth pulse (15 ns to 5s)
SQUARE
Square wave (50% duty cycle) (4 ns to 5s)
SQ20PCT
Square wave (20% duty cycle) (85 ns to 5s)
Example:
TMWAVE SPIKE
TMWAVE?
(IEEE-488, RS-232, Sequential)
Shows the MARKER mode waveform setting.
Parameters: None
Response: <character> (Shows
SINE, SPIKE, SQUARE
, or
SQ20PCT
.)
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...