SC300 Oscilloscope Calibration Option
Verification Tables
10
10-31
Table 10-4. Voltage Function Verification: DC Voltage into a 1 M
Ω
Load
Nominal Value
(dc)
Measured Value (dc)
Deviation (mV)
1-Year Spec.
0.0 mV
0.10 mV
5.0 mV
0.11 mV
-5.0 mV
0.11 mV
22.0 mV
0.15 mV
-22.0 mV
0.15 mV
25.0 mV
0.16 mV
-25.0 mV
0.16 mV
45.0 mV
0.21 mV
-45.0 mV
0.21 mV
50.0 mV
0.23 mV
-50.0 mV
0.23 mV
220.0 mV
0.65 mV
-220.0 mV
0.65 mV
250.0 mV
0.72 mV
-250.0 mV
0.72 mV
450.0 mV
1.22 mV
-450.0 mV
1.22 mV
500.0 mV
1.35 mV
-500.0 mV
1.35 mV
3.3 V
8.35 mV
-3.3 V
8.35 mV
4.0 V
10.10 mV
-4.0 V
10.10 mV
33.0 V
82.60 mV
-33.0 V
82.60 mV