SC600 Oscilloscope Calibration Option
Oscilloscope Connections
9
9-7
Overload Measurement Specifications
Source Voltage
Typical ‘On’ Current
Indication
Typical ‘Off’ Current Indication
Maximum Time Limit DC or AC
(1 kHz)
5 to 9 V
100 to 180 mA
10 mA
Setable 1 s to 60 s
Oscilloscope Connections
With the cable supplied with the SC600 Option, connect the SCOPE output on the
Calibrator to one of the channel connectors on your oscilloscope (see Figure 9-1).
To use the external trigger, connect the TRIG output on the Calibrator to the external
trigger connection on your oscilloscope. To use the external trigger and see its signal with
the calibration signal, connect the TRIG output to A different channel. See your
oscilloscope manual for instructions to connect and see an external trigger.
5502A CALIBRATOR
gvx036.eps
Figure 9-1. Oscilloscope Connection: Channel and External Trigger
Start the SC600 Option
Push
(LED on) to select the SC600 Option. The SCOPE menu, shown below,
shows in the Control Display. You can push the first four softkeys to go directly to the
VOLT, EDGE, LEVSINE, and MARKER calibration menus. Push the last softkey to go
to the OTHER menu (also shown below), this lets you access to WAVEGEN, VIDEO,
PULSE, Impedance/Capacitance measurement (MEAS Z), and Overload (OVERLD)
menus. Push
to return to the SCOPE menu from the OTHER menu. This chapter
tells you about each of these menus.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...