5502A
Operators Manual
7-18
Table 7-11. Verification Tests for Thermocouple Simulation
TC Type
Output,
°
C
Lower Limit, mV
Upper Limit, mV
10
μ
V/
°
C
0.00
°
C (0.0000 mV)
-0.0030
0.0030
100.00
°
C (1.0000 mV)
0.99696
1.00304
-100.00
°
C (-1.0000 mV)
-1.00304
-0.99696
1000.00
°
C (10.0000 mV)
9.99660
10.00340
-1000.00
°
C (10.0000 mV)
-10.0034
-9.9966
10000.00
°
C (100.0000 mV)
99.9930
100.0070
-10000.00
°
C (-100.0000 mV) -100.0070
-99.9930
Table 7-12. Verification Tests for Thermocouple Measurement
TC Type
Input, mV
Lower Limit,
°
C Upper
Limit,
°
C
10
μ
V/
°
C
0.00
°
C (0,0000 mV)
-0.30 -0.30
10000.00
°
C (100.0000 mV)
9999.30
10000.70
-10000.00
°
C (-100.0000 mV) -10000.70
-9999.30
30000.00
°
C (300.0000 mV)
29998.50
30001.50
-30000.00
°
C (-300.0000 mV) -30001.50
-29998.50