5502A
Operators Manual
4-28
Note
At voltage outputs of 100 V and above (nominal), the Calibrator usually
makes a small high-pitched sound.
5.
Push a multiplier key, if necessary. For example, push
.
6.
Push
.
7.
The Control Display shows the amplitude of your voltage entry. For example,
123.456 mV (see below).
nn084f.eps
8.
Push the numeric keys and decimal point key to record the necessary current output
(maximum six numeric keys). For example, 234.567.
9.
Push a multiplier key, if necessary. For example, push
.
10.
Push
.
11.
The Control Display shows the amplitude of your voltage and current entries. For
example, 123.456 mV and 234.567 mA (see below).
nn085f.eps
12.
Push the numeric keys and decimal point key to record the necessary frequency
output (maximum five numeric keys). Push a multiplier key, if necessary. For
example, push the kilo multiplier key
. Then push the
key. For example,
1.1234 kHz.
13.
The Control Display shows your entries. For example, 123.456 mV and 234.567 mA
at 1.1234 kHz (see below).
nn086f.eps
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...