9-4
SC600 Oscilloscope Calibration Option Specifications
These specifications apply only to the SC600 Oscilloscope Calibration Option. General
specifications that apply to the 5502A Calibrator can be found in Chapter 1. The
specifications are correct given the 5502A is operated with the conditions specified in
Chapter 1, and has completed a warm-up period of no less than two times the length of
time the Calibrator was powered off, to a maximum of 30 minutes.
Voltage Function Specifications
Voltage Function
DC Signal
Square Wave Signal
[1]
50
Ω
Load
1 M
Ω
Load
50
Ω
Load
1 M
Ω
Load
Amplitude Qualities
Range
0 to
±
6.599 V
0 to
±
130 V
±
1 mV to
±
6.599 V p-p
±
1 mV to
±
130 V p-p
Resolution
Range
1 to 24.999 mV
25 to 109.99 mV
110 mV to 2.1999 V
2.2 to 10.999 V
11 to 130 V
Resolution
1
μ
V
10
μ
V
100
μ
V
1 mV
10 mV
Adjustment Range
Continuously adjustable
1-Year Absolute Uncertainty,
tcal
±
5
°
C
±
(0.25 % of output
+ 40
μ
V)
±
0.05 % of output
+ 40
μ
V)
±
(0.25 % of output
+ 40
μ
V)
±
(0.1 % of
40
μ
V)
[2]
Sequence
1-2-5 (e.g., 10 mV, 20 mV, 50 mV)
Square Wave Frequency Qualities
Range
10 Hz to 10 kHz
1-Year Absolute Uncertainty,
tcal
±
5
°
C
±
(2.5 ppm of setting)
Typical aberration within 4
μ
s from
50 % of leading/trailing edge
<(0.5 % of 100
μ
V)
[1]
Selectable positive or negative, zero referenced square wave.
[2]
For square wave frequencies above 1 kHz,
±
(0.25 % of 40
μ
V).
Edge Specifications
Edge Qualities into 50
Ω
Load
1-Year Absolute Uncertainty,
tcal
±
5
°
C
Rise Time
≤
300 ps
(+0 ps / -100 ps)
Amplitude Range (p-p)
4.5 mV to 2.75 V
±
(2 % of 200
μ
V)
Resolution 4
digits
Adjustment Range
±
10 % around each sequence value
(indicated below)
Sequence Values
5 mV, 10 mV, 25 mV, 50 mV, 60 mV, 80
mV, 100 mV, 200 mV, 250 mV, 300 mV,
500 mV, 600 mV, 1 V, 2.5 V
Frequency Range
[1]
900 Hz to 11 MHz
±
(2.5 ppm of setting)
Typical Jitter, edge to trigger
<5 ps (p-p)
Leading Edge Aberrations
[2]
within 2 ns from 50 % of rising edge
<(3 % of 2 mV)
2 to 5 ns
<(2 % of 2 mV)
5 to 15 ns
<(1 % of 2 mV)
after 15 ns
<(0.5 % of 2 mV)
Typical Duty Cycle
45 % to 55 %
Tunnel Diode Pulse Drive
Square wave at 100 Hz to 100 kHz, with variable amplitude of 60 to 100 V p-p.
[1]
Above 2 MHz rise time specification <350 ps
[2]
All edge aberration measurements made with Tektronix 11801 mainframe with SD26 input module.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...