SC300 Oscilloscope Calibration Option
Calibrate the Time Base of an Oscilloscope
10
10-23
Time Base Marker Calibration Procedure for an Oscilloscope
This sample procedure uses the Time Marker function to verify the horizontal deflection
(time base) of your oscilloscope. See your manual for the oscilloscope for the time base
values recommended for calibration.
Before you start this procedure, make sure that you are in Marker mode. If you are, the
Control Display shows the subsequent menu, see below.
gl034i.eps
Do the subsequent sample procedure to calibrate the time base.
1.
Connect the Calibrator to Channel 1 on the oscilloscope. Select 50
Ω
impedance or
use an external 50
Ω
termination. Make sure the oscilloscope is dc-coupled.
2.
Apply a time marker value. Refer to the recommended calibration settings in your
oscilloscope manual. For example, to key in 200 ns, push
, and then push
.
Note
You can record the equivalent frequency as an alternative to the time
marker value. For example, as an alternative to 200 ns, you can record
5 MHz.
3.
Set the time base of your oscilloscope to show 10 time markers. The time markers
must align with the oscilloscope divisions, as shown in the example below. For an
accurate indication, align the peaks of the signal with the horizontal center axis, see
below.
Peaks are aligned
w
ith center axis
gl011i.eps
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...