5502A
Operators Manual
9-42
Table 9-8. SC600 Option Leveled Sine Wave Verification: Frequency
Nominal Value
(V p-p)
Frequency
Measured Value
(Hz)
Deviation (Hz)
1-Year Spec. (Hz)
5.5 50
kHz
0.125
5.5 500
kHz
1.25
5.5 5
MHz
12.5
5.5 50
MHz
125
5.5 500
MHz
1250
Table 9-9. SC600 Option Leveled Sine Wave Verification: Harmonics
Harmonic
Nominal Value
(V p-p)
Frequency
Measured
Value (dB)
Deviation
(dB)
1-Year Spec.
(dB)
2nd harmonic
0.0399
50 kHz
-33
3rd+ harmonic
0.0399
50 kHz
-38
2nd harmonic
0.099
50 kHz
-33
3rd+ harmonic
0.099
50 kHz
-38
2nd harmonic
0.399
50 kHz
-33
3rd+ harmonic
0.399
50 kHz
-38
2nd harmonic
1.2
50 kHz
-33
3rd+ harmonic
1.2
50 kHz
-38
2nd harmonic
5.5
50 kHz
-33
3rd+ harmonic
5.5
50 kHz
-38
2nd harmonic
5.5
100 kHz
-33
3rd+ harmonic
5.5
100 kHz
-38
2nd harmonic
5.5
200 kHz
-33
3rd+ harmonic
5.5
200 kHz
-38
2nd harmonic
5.5
400 kHz
-33
3rd+ harmonic
5.5
400 kHz
-38
2nd harmonic
5.5
800 kHz
-33
3rd+ harmonic
5.5
800 kHz
-38
2nd harmonic
5.5
1 MHz
-33
3rd+ harmonic
5.5
1 MHz
-38
2nd harmonic
5.5
2 MHz
-33
3rd+ harmonic
5.5
2 MHz
-38
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...