Introduction
and
Specifications
Detailed Specifications
1
1-17
Capacitance
Range
Absolute Uncertainty, tcal
±
5
°
C
±
(% of floor)
[1] [2] [3]
Resolution
Allowed Frequency or Charge-Discharge Rate
90 Day
1 Year
Min and Max to
Meet
Specification
Typical Max for
<0.5 % Error
Typical Max for
<1 % Error
220.0 to 399.9 pF 0.38 + 0.01 nF
0.5 + 0.01 nF
0.1 pF
10 Hz to 10 kHz
20 kHz
40 kHz
0.4 to 1.0999 nF 0.38 + 0.01 nF
0.5 + 0.01 nF
0.1 pF
10 Hz to 10 kHz
30 kHz
50 kHz
1.1 to 3.2999 nF 0.38 + 0.01 nF
0.5 + 0.01 nF
0.1 pF
10 Hz to 3 kHz
30 kHz
50 kHz
3.3 to 10.999 nF 0.19 + 0.01 nF 0.25 + 0.01 nF
1 pF
10 Hz to 1 kHz
20 kHz
25 kHz
11 to 32.999 nF
0.19 + 0.1 nF
0.25 + 0.1 nF
1 pF
10 Hz to 1 kHz
8 kHz
10 kHz
33 to 109.99 nF
0.19 + 0.1 nF
0.25 + 0.1 nF
10 pF
10 Hz to 1 kHz
4 kHz
6 kHz
110 to 329.99 nF 0.19 + 0.3 nF
0.25 + 0.3 nF
10 pF
10 Hz to 1 kHz
2.5 kHz
3.5 kHz
0.33 to
1.0999
μ
F
0.19 + 1 nF
0.25 + 1 nF
100 pF
10 to 600 Hz
1.5 kHz
2 kHz
1.1 to 3.2999
μ
F
0.19 + 3 nF
0.25 + 3 nF
100 pF
10 to 300 Hz
800 Hz
1 kHz
3.3 to 10.999
μ
F
0.19 + 10 nF
0.25 + 10 nF
1 nF
10 to 150 Hz
450 Hz
650 Hz
11 to 32.999
μ
F
0.30 + 30 nF
0.40 + 30 nF
1 nF
10 to 120 Hz
250 Hz
350 Hz
33 to 109.99
μ
F 0.34 + 100 nF
0.45 + 100 nF
10 nF
10 to 80 Hz
150 Hz
200 Hz
110 to 329.99
μ
F 0.34 + 300 nF
0.45 + 300 nF
10 nF
0 to 50 Hz
80 Hz
120 Hz
0.33 to
1.0999 mF
0.34 + 1
μ
F
0.45 + 1
μ
F
100 nF
0 to 20 Hz
45 Hz
65 Hz
1.1 to 3.2999 mF
0.34 + 3
μ
F
0.45 + 3
μ
F
100 nF
0 to 6 Hz
30 Hz
40 Hz
3.3 to 10.999 mF 0.34 + 10
μ
F
0.45 + 10
μ
F 1
μ
F
0 to 2 Hz
15 Hz
20 Hz
11 to 32.999 mF
0.7 + 30
μ
F
0.75 + 30
μ
F 1
μ
F
0 to 0.6 Hz
7.5 Hz
10 Hz
33 to 110.00 mF
1.0 + 100
μ
F
1.1 + 100
μ
F 10
μ
F
0 to 0.2 Hz
3 Hz
5 Hz
[1]
The output is continuously variable from 220 pF to 110 mF.
[2]
Specifications apply to both dc charge/discharge capacitance meters and ac RCL meters. The maximum allowable peak voltage is
3 V. The maximum allowable peak current is 150 mA, with an rms limitation of 30 mA below 1.1
μ
F and 100 mA for 1.1
μ
F and
above.
[3]
The maximum lead resistance for no additional error in 2-wire COMP mode is 10
Ω
.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...