5502A
Operators Manual
3-22
Table 3-3. Factory Defaults for SETUP Menus Power-Up Defaults
Parameter Setting
SETUP Menu
(Figure 3-4)
User report string (*PUD string)
Cleared
D
Error units
>
0.1%
G1
SC-600 option overload test safety
timeout
10 s
G1
Temperature standard
its-90
G1
Host interface
gpib (IEEE-488)
G1
UUT serial interface
8 bits, 1 stop bit, xon/xoff, parity
none, 9600 baud
I
Host serial interface
term, 8 bits, 1 stop bit, xon/xoff,
parity none, 9600 baud, CRLF,
012,000
K, L, M, N
GPIB Port Address
4
O
Display brightness (Note)
level 1,0
P
Display contrast (Note)
level 7,7
P
dBm impedance
600
Ω
S
RTD type
pt385
S1
Thermocouple type
K
S1
Current limits
±
20.5 A
U
Voltage limits
±
1020 V
V
Note:
Output Display and Control Display, respectively. There are 8 levels: 0,1,2,3,4,5,6, and 7