Remote Commands
Commands
6
6-41
UNCERT?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Uncertainties command) Shows specified uncertainties for the current output. If there are
no specifications for an output, shows zero.
Parameter: 1. (optional) Preferred unit of primary output uncertainty or PCT (default)
2. (optional) Preferred unit of secondary output uncertainty or PCT
(default)
Response:
1. (float) 90-day specified uncertainty of primary unit
2. (float) 1-year specified uncertainty of primary output
3. (character) Unit of primary output uncertainty
4. (float) 90-day specified uncertainty of secondary unit
5. (float) 1-year specified uncertainty of secondary output
6. (character) Unit of secondary output uncertainty.
Example:
UNCERT?
shows
6.120E-01,6.150E-01,PCT,9.50E-02,
1.150E-01,PCT
UUT_FLUSH
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Flush UUT Receive Buffer command) Flushes the UUT receive buffer for data received
from the UUT over the Calibrator rear panel SERIAL 2 TO UUT serial port. The
command can be sent over GPIB or RS-232 ports, but applies to SERIAL 2 TO UUT
serial port operation.
Parameter: (None)
Example:
UUT_FLUSH
Flush the Calibrator receive data buffer for the UUT.
UUT_RECV?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(UUT Receive Data query) Shows data from the UUT in IEEE-488.2 Standard format
over the Calibrator rear panel SERIAL 2 TO UUT serial port. The command can be sent
over GPIB or RS-232 ports, but applies to SERIAL 2 TO UUT serial port operation.
Response:
<data> (binary block data in definite length format from UUT)
Example:
UUT_RECV?
shows
#211+1.0
Shows (for example) a measurement from the UUT. The format is
#2
(two numbers
follow)
11
(characters follow)
+1.0
(11 characters).
UUT_RECVB?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(UUT Receive Binary Data query) Shows binary data as integers from the UUT serial
port. Use the UUT_RECV? command instead if receiving ASCII data.
Parameter: (Optional) Maximum number of integers per line
Response:
(Indefinite ASCII) Comma separated integers as follows:
1. (integer) Number of data bytes shown that do not include the count
2. (integer) Data from the UUT serial port as series of comma separated
integers
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...