Front-Panel Operation
Set Output
4
4-19
Set DC Voltage Output
Do these procedures to set a dc voltage output from the front-panel NORMAL terminals.
If you make an entry error, push
to clear the display, then record the value again.
Caution
To prevent damage to the UUT, make sure the applied voltage
to the UUT is not larger than the rating of the UUT insulation
and the connection wires.
1.
Push
to clear outputs from the Calibrator.
2.
Connect the UUT as in the “Connect the Calibrator to a UUT” section of this chapter.
3.
Set the UUT to measure dc voltage on the necessary range.
4.
Push the numeric keys and decimal point key to record the necessary voltage output
(maximum seven numeric keys). For example, 123.4567.
Note
At voltage outputs of 100 V and above (nominal), the Calibrator usually
makes a small high-pitched sound.
5.
Push
to select the polarity of the voltage (default is +)
6.
Push a multiplier key, if necessary. For example, push
.
7.
Push
.
8.
The Control Display shows the amplitude of your entry. For example,
123.4567 mV (see below).
nn071f.eps
9.
Push
. The Calibrator erases your entry from the Control Display and copies
it into the Output Display (below is typical).
nn072f.eps
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...