5502A
Operators Manual
6-46
ZERO_MEAS
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
Sets the zero offset for capacitance measurement with the -SC600.
Parameter:
1. (boolean) ON
(boolean) OFF
Example:
ZERO_MEAS ON
Sets the zero offset to the current measurement value.
ZERO_MEAS?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
Shows the zero offset for capacitance measurements with the -SC600.
Parameter:
(None)
Responses: 1. (character) OFF
(no zero in effect)
(character) ON
(zero in effect)
2. (float) offset value
3. (character) units
(F)
Example:
ZERO_MEAS?
shows
ON,-3.66E-02,F