Remote Commands
Commands
6
6-33
Shows
PT3926
when the RTD default type is a 100-ohm RTD with curve
α
=0.003926 ohms/ohm/
°
C.
SP_SET
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Host Serial Port Set command) Sets the RS-232-C settings for the Calibrator rear panel
SERIAL 1 FROM HOST serial port, which is kept in the Calibrator non-volatile
memory. (While saving configuration data in the non-volatile memory, a period of about
2 seconds, the Calibrator does not respond to remote commands.) (To set the parameters
for the rear panel SERIAL 2 TO UUT serial port, see the
UUT_SET
command.) The
factory default values are shown below in
BOLD
type. (To return to the factory defaults,
see the
FORMAT SETUP
command.)
The interface selection sets the command response, with command echo back for
commands and error messages with TERM (terminal) or no echo back with COMP
(computer).
Parameters: <baud rate value>,
300, 600, 1200, 2400, 4800,
9600
<interface>,
TERM
(terminal), COMP (computer)
<flow
control>,
XON
(xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>,
DBIT7 (7 bits) or
DBIT8
(8 bits)
<number
stop
bits>,
SBIT1
(1 bit) or SBIT2 (2 bits)
<parity>,
PNONE
(none), PODD (odd),PEVEN (even)
<end of line char.>
CR (carriage return), LF (line feed),
CRLF
(carriage return/line feed)
Example:
SP_SET 9600,TERM,XON,DBIT8,SBIT1,PNONE,CRLF
Set the parameters for the rear panel SERIAL 1 FROM HOST serial port to the factory
default values.
SP_SET?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Host Serial Port Set query) Shows the RS-232-C settings for the Calibrator rear panel
SERIAL 1 FROM HOST serial port. (To return the parameters for the rear panel
SERIAL 2 TO UUT serial port, see the
UUT_SET?
command.) The factory default
values are shown below in
BOLD
type. (To return to the factory defaults, see the
FORMAT SETUP
command.)
Responses: <baud rate value>,
300, 600, 1200, 2400, 4800,
9600
<interface>,
TERM
(terminal), COMP (computer)
<flow
control>,
XON
(xon/xoff), NOSTALL (none), RTS (rts/cts)
<number data bits>,
DBIT7 (7 bits) or
DBIT8
(8 bits)
<number
stop
bits>,
SBIT1
(1 bit) or SBIT2 (2 bits)
<parity>,
PNONE
(none), PODD (odd),PEVEN (even)
<end of line char.>
CR (carriage return), LF (line feed),
CRLF
(carriage return/line feed)
Example:
SP_SET?
shows
9600,TERM,XON,DBIT8,SBIT1,PNONE,CRLF
Shows the parameters for the rear panel SERIAL 1 FROM HOST serial port, as shown,
when set to the factory default values.
SPLSTR
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Serial Poll String command) Sets the Serial Poll String (string up to 40 characters)
which is kept in the Calibrator non-volatile memory. (While saving configuration data in
the non-volatile memory, a period of about 2 seconds, the Calibrator does not respond to
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...