Remote Operation
Use Commands
5
5-31
before they have completed execution.
In Chapter 6, the command shows a checkbox for overlapped commands.
The overlapped commands, but not the scope commands, are:
DBM OPER
TSENS_TYPE
DC_OFFSET OUT
WAVE
DPF PHASE
ZCOMP
DUTY RANGELCK
EARTH *RST
HARMONIC RTD_TYPE
INCR STBY
LCOMP TC_OFFSET
LOWS TC_OTCD
MULT TC_REF
OLDREF TC_TYPE
You can use the command
*WAI
to wait until the overlapped command has completed
execution before you do the subsequent command. For example,
OUT 1 V, 1 A, 60 HZ; *WAI
You can also use the status commands
*OPC
and
*OPC?
to find completion of
overlapped commands. (See “Check 5502A Status.”)
Sequential Commands
Commands that start immediately are sequential commands.
In Chapter 6, the command shows a checkbox for sequential commands.
Most of the commands are sequential.
Commands where a Calibration Switch is Necessary
For the subsequent commands, the rear panel CALIBRATION switch must be in the
ENABLE position:
CLOCK
(when setting date but not time)
FORMAT ALL
FORMAT CAL
*PUD
If you try to use these commands with the CALIBRATION switch in the NORMAL
position, an error is logged into the error queue. (Or it returns the error message if in the
RS-232 Terminal Mode.)
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...