5502A
Operators Manual
1-28
AC Current (Non-Sine Wave) (cont.)
Square Wave Range p-p
Frequency
1-Year Absolute Uncertainty tcal
±
5
°
C
±
(% of % of range)
Max Current
Resolution
0.047 to 0.65999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
10 + 2
0.66 to 6.59999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
10 + 2
6.6 to 65.9999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.25
1 to 10 kHz
10 + 2
66 to 659.999 mA
[1]
10 to 45 Hz
0.25 + 0.5
Six digits
45 Hz to 1 kHz
0.25 + 0.5
1 to 10 kHz
10 + 2
0.66 to 5.99999 A
[2]
10 to 45 Hz
0.5 + 1.0
45 Hz to 1 kHz
0.5 + 0.5
1 to 10 kHz
10 + 2
6 to 41 A
[2]
45 to 500 Hz
0.5 + 0.5
500 Hz to 1 kHz
1.0 + 1.0
[1]
Frequency limited to 1 kHz with LCOMP on.
[2]
Frequency limited to 440 Hz with LCOMP on.
AC Current, Square Wave Characteristics (typical)
Range LCOMP
Risetime Settling
Time
Overshoot
I <6 A @ 400 Hz
off
25
μ
s 40
μ
s to 1 % of final value
<10 % for <1 V Compliance
3 A & 20 A Ranges
on
100
μ
s 200
μ
s to 1 % of final value
<10 % for <1 V Compliance
AC Current, Triangle Wave Characteristics (typical)
Linearity to 400 Hz
Aberrations
0.3 % of p-p value, from 10 % to 90 % point
<1 % of p-p value, with amplitude >50 % of range
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...