Remote Commands
Commands
6
6-19
Responses: (Indefinite ASCII) A message that has four fields separated by commas as
follows:
1.
Manufacturer
2.
Model
number
3. Serial number
4. Firmware revision levels for the Main CPU
+
Front Panel CPU
+
Inguard
PGA
Example:
*IDN?
shows
FLUKE,5502A,5248000,1.0+1.3+1.8
Shows Fluke manufacturer, model 5502A, serial number 5248000, main firmware
version 1.0, encoder firmware 1.3, and inguard PGA 1.8.
INCR
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
(Increment command) Increments or decrements the output (as selected if you use the
EDIT
command, or defaults to the primary output) and records error mode; the same as if
you use the Calibrator output adjustment knob in local operation.
Parameters: <+ value>
(increment value) (optional unit matching edit field)
<–value>
(decrement
value)
Example:
INCR +.00001 mV
Load the error mode and increment the selected edit field by .00001 mV.
ISCE
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Instrument Status Change Enable command) Puts two bytes into the two 16-bit ISCE
mask registers (ISCE1 and ISCE0). (See “Instrument Status Change Enable Registers” in
Chapter 5 for more information.)
Parameter: <value>
(decimal
equivalent of the 16 bits, 0 to 32767)
Example:
ISCE 6272
Load decimal 6272 (binary
0001010001000000
) to enable bits 12 (SETTLED), 10
(REMOTE) and 6 (HIVOLT). This is equivalent to sending the commands
ISCE0 6272
and
ISCE1 6272
(see below).
ISCE?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Instrument Status Change Enable query) Shows the two bytes from the two 16-bit ISCE
mask registers (ISCE1 and ISCE0). (See “Instrument Status Change Enable Registers” in
Chapter 5 for more information.)
Response:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCE?
shows
6272
Shows decimal
6272
(binary
0001010001000000
) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
ISCE0
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
(Instrument Status 1 to 0 Change Enable command) Puts the two bytes into the 16-bit
ISCE0 register. (See “Instrument Status Change Enable Registers” in Chapter 5 for more
information.)
Parameter: <value>
(decimal
equivalent of the 16 bits, 0 to 32767)
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...