5502A
Operators Manual
1-10
outputs (e.g. 1000 V, 20 A) choose a ling voltage
setting that is
±
7.5 % from nominal.
Power Consumption
............................................. 600 VA
Dimensions (HxWxL) .............................................. 17.8 cm x 43.2 cm x 47.3 cm (7 in x 17 in x 18.6 in) Standard rack
width and rack increment, plus 1.5 cm (0.6 in) for feet on bottom of
unit.
Weight (without options) ......................................... 22 kg (49 lb)
Absolute Uncertainty Definition
.......................... The 5502A specifications include stability, temperature coefficient,
linearity, line and load regulation, and the traceability of the external
standards used for calibration. You do not need to add anything to
determine the total specification of the 5502A for the temperature
range indicated.
Specification Confidence Level
........................... 99 %
Detailed Specifications
DC Voltage
Range
Absolute Uncertainty, tcal ± 5
°
C
±(% of
μ
V)
Stability
Resolution (µV)
Max Burden
[1]
24 hours, ± 1
°
C
±(ppm of
μ
V)
90 Day
1 Year
0 to 329.9999 mV
0.005 + 3
0.006 + 3
5 + 1
0.1 65
Ω
0 to 3.299999 V
0.004 + 5
0.005 + 5
4 + 3
1 10
mA
0 to 32.99999 V
0.004 + 50
0.005 + 50
4 + 30
10 10
mA
30 to 329.9999 V
0.0045 + 500
0.0055 + 500
4.5 + 300
100 5
mA
100 to 1020.000 V
0.0045 + 1500
0.0055 + 1500
4.5 + 900
1000 5
mA
Auxiliary Output (dual output mode only)
[2]
0 to 329.999 mV
0.03 + 350
0.04 + 350
30 + 100
1 5
mA
0.33 to 3.29999 V
0.03 + 350
0.04 + 350
30 + 100
10 5
mA
3.3 to 7 V
0.03 + 350
0.04 + 350
30 + 100
100 5
mA
TC Simulate and Measure in Linear 10
μ
V/°C and 1 mV/
°
C modes
[3]
0 to 329.999 mV
0.005 + 3
0.006 + 3
5 + 1
0.1 10
Ω
[1]
Remote sensing is not provided. Output resistance is < 5 m
Ω
for outputs
≥
0.33 V. The AUX output has an output resistance of
<1
Ω
. TC simulation has an output impedance of 10
Ω
± 1
Ω
.
[2]
Two channels of dc voltage output are provided.
[3]
TC simulating and measuring are not specified for operation in electromagnetic fields above 0.4 V/m.
Range
Noise
Bandwidth 0.1 Hz to 10 Hz p-p
±
(ppm of floor in
μ
V)
Bandwidth 10 Hz to 10 kHz rms
0 to 329.9999 mV
0 + 1
6
μ
V
0 to 3. 3.299999 V
0 + 10
60
μ
V
0 to 32.99999 V
0 + 100
600
μ
V
30 to 329.9999 V
10 + 1000
20 mV
100 to 1020.000 V
10 + 5000
20 mV
Auxiliary Output (dual output mode only)
[1]
0 to 329.999 mV
0 + 5
μ
V 20
μ
V
0.33 to 3.29999 V
0 + 20
μ
V 300
μ
V
3.3 to 7 V
0 + 100
μ
V 1000
μ
V
[1]
Two channels of dc voltage output are provided.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...