![Fluke 5502A Operator'S Manual Download Page 156](http://html1.mh-extra.com/html/fluke/5502a/5502a_operators-manual_2310344156.webp)
5502A
Operators Manual
5-14
The characters REMOTE must have shown on the terminal screen as they were
recorded. If they did not show on the screen, but the Control Display changed to
REMOTE CONTROL, then see step 4 of the “RS-232 Host Port Setup Procedure”
and change the REMOTE I/F setting from comp to term.
If nonsense characters showed on the screen, then there is a mismatch of RS-232
parameters. See step 4 of the “RS-232 Host Port Setup Procedure” procedure for the
correct RS-232 settings and then do this procedure again. Start at Step 5.
If no characters showed on the screen, then see step 3 of the “RS-232 Host Port
Setup Procedure” procedure to make sure that serial was selected for the Host port.
Check that you used the correct RS-232 cable. It must be in a null-modem
configuration where the RX and TX lines are reversed (see Appendix C). Also make
sure that you have connected to the correct COM port on the PC.
8.
Type the command LOCAL and push <Enter>. See that the Calibrator Control
Display changes to the reset condition (below).
nn323f.eps
To experiment with other commands in the command set, see Chapter 6, “Remote
Commands.” When finished, select the Exit command from the File menu to close
the Terminal accessory.
Hint: To keep the communication parameters in Terminal for future operations, first
select Save from the File menu and then give a name, for example, host.trm.
Verify RS-232 Host Port Operation with Visual Basic
Complete the subsequent procedure to verify RS-232 (Host) operation with the
Windows-based programming language Visual Basic. This procedure is if you have
completed Appendix D, “Creating a Visual Basic Test Program” to make the group
RS-232 Test.
Complete the subsequent procedure to verify RS-232 operation with Visual Basic.
1.
Complete the “RS-232 Host Port Setup Procedure” before in this chapter to set up the
Calibrator for RS-232 Host port operation. Note the RS-232 Host port parameters
that you selected in this procedure.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...