SC600 Oscilloscope Calibration Option
Test Video Triggers
9
9-25
Test Video Triggers
gjh062.eps
You can push the MODE softkey to move through the functions in the sequence shown,
or you can push
to return directly to the OTHER modes menu.
Each option in the Video menu is given below.
•
OUTPUT @ SCOPE terminal (50
Ω
)
Shows the location of the signal output. If
the signal does not show on the oscilloscope, push
. To disconnect the signal,
push
.
•
LINE MK
Lets you to select the marker line number. For ntsc and pal-m formats,
you can also select field (“odd” or “even”). For pal and secam formats, the field
(“ODD” or “EVEN”) is selected automatically in reference to the marker line
number.
•
FORMAT
Scrolls through the available formats. You can select ntsc, pal, pal-m,
and secam.
•
MODE
Shows you are in VIDEO mode. Use the softkey to change modes and open
the related menus for the other four oscilloscope calibration modes.
Default video settings are +100 %, format = NTSC, and videomark = 10.