![Fluke 5502A Operator'S Manual Download Page 94](http://html1.mh-extra.com/html/fluke/5502a/5502a_operators-manual_2310344094.webp)
5502A
Operators Manual
4-20
10.
Push
to turn on the Calibrator output.
A softkey name for range is shown on the Control Display in the dc voltage function:
nn063f.eps
Range
(Operating Range) selects autorange (auto) or lock (locked) for the current
range. When auto (the default) is selected, the Calibrator automatically selects the
range that gives the best output resolution. When locked is selected, the Calibrator
will not change ranges when you edit the output. The locked selection is typically
made to prevent range changes that can cause a small agitation in the output, for
example, when you measure the linearity of a given multimeter range.
Set AC Voltage Output
You can select an ac voltage output in volts or as a power output in dBm, where dBm is
10 log (Pout/.001), where Pout is shown in watts. The output range is 1 mV to 1000 V.
When you select dBm outputs, the Calibrator calculates dBm at a selected impedance
level. For this instance, the formula is:
20 log(V) - 10 log(Impedance * .001) = dBm.
Do this procedure to set an ac voltage output at the front panel NORMAL terminals. If
you make an entry error, push
to erase the display, then record the value again.
Caution
To prevent damage to the UUT, makes sure the applied voltage to
the UUT is not larger than the rating of the UUT insulation and the
connection wires.
1.
Push
to clear output from the Calibrator.
1.
Connect the UUT as in the “Connect the Calibrator to a UUT” section of this chapter.
2.
Set the UUT to measure ac voltage on the necessary range.
Output in volts:
Push the numeric keys and decimal point key to record the necessary voltage
output (maximum six numeric keys). For example, 2.44949.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...