Remote Commands
6
6-9
Table 6-9. Status Commands
Command Description
ERR?
Shows the first error code with an explanation in the Calibrator error queue, and then
removes that error code from the queue.
EXPLAIN?
Tells you about an error code. This command shows a string that tells you about the
error code given as the parameter.
FAULT?
Shows the first error code contained in the Calibrator error queue, and then removes that
error from the queue.
FUNC?
Shows the current output, measurement, or calibration function.
ISCE
Puts two bytes into both the Instrument Status 1 to 0 Change Enable register and the
Instrument Status 0 to 1 Change Enable register.
ISCE?
Shows the OR of the contents of the Instrument Status 1 to 0 Change Enable register
and the Instrument Status 0 to 1 Change Enable register.
ISCE0
Puts two bytes into the Instrument Status 1 to 0 Change Enable register.
ISCE0?
Shows the contents of the Instrument Status 1 to 0 Change Enable register.
ISCE1
Puts two bytes into the Instrument Status 0 to 1 Change Enable register.
ISCE1?
Shows the contents of the Instrument Status 0 to 1 Change Enable register.
ISCR?
Shows the OR of the contents of the Instrument Status 1 to 0 Change Register and the
Instrument Status 0 to 1 Change Register and clears the registers.
ISCR0?
Shows and clears the contents of the Instrument Status 1 to 0 Change Register.
ISCR1?
Shows and clears the contents of the Instrument Status 0 to 1 Change Register.
ISR?
Shows the contents of the Instrument Status Register.
Table 6-10. Thermocouple (TC) Measurement Commands
Command Description
TC_MEAS
Changes the operation mode to thermocouple measurement.
TC_OFFSET
Sets a temperature offset for the thermocouple measurement mode.
TC_OFFSET?
Shows the temperature offset when in the thermocouple measurement mode.
TC_OTCD?
Shows if the open thermocouple detection circuit is set.
TC_OTCD
Activates or deactivates the open thermocouple detection circuit in thermocouple
measurement mode.
VAL?
Shows the last thermocouple value, or, for the -SC600, impedance measurement value.
VVAL?
Shows the last value of the thermocouple measurement in volts.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...