SC600 Oscilloscope Calibration Option
Test the Trigger SC600 Option
9
9-23
Note
You can record the equivalent frequency as an alternative to the time
marker value. For example, you can record 5 MHz for 200 ns.
3.
Set the time base of your oscilloscope to show 10 time markers. The time markers
must align with the oscilloscope divisions, as shown in the example below.
For an accurate indication, align the peaks of the signal with the horizontal center
axis.
Peaks are aligned
w
ith center axis
gl011i.eps
4.
Do this procedure again for all time marker values recommended for your
oscilloscope. Do this procedure again for digital and analog mode as necessary. It can
be necessary, for some oscilloscopes, to change the magnification while you calibrate
in analog mode.
5.
Push
to remove the input signal.
Test the Trigger SC600 Option
The oscilloscope can trigger on different waveforms. This can be tested with the wave
generator. When the wave generator is used, a square, sine, or triangle wave is
transmitted and the wave output impedance, offset, and voltage can be changed to test the
trigger function at different levels.
Note
The wave generator must not be used to verify the accuracy of your
oscilloscope.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...