5502A
Operators Manual
4-56
5502A CALIBRATOR
87
TRUE RMS MULTIMETER
MIN MAX
RANGE
HOLD
H
REL
Hz
mA
A
A
m
mV
V
V
OFF
PEAK MI
N
MAX
A
COM
V
mA A
gvx025.eps
Figure 4-15. Cable Connections to Verify 80 Series General Functions
3.
Verify the dc voltage function:
a.
Turn on the DMM and set the function switch to
.
b.
Set the warmed-up Calibrator to 3.5 V dc.
c.
Push
.
d.
Use the output-adjustment controls to adjust the Calibrator output to measure
+3.5000 on the DMM display.
e.
Make sure that the error shown on the control display is below the specification
that is in the DMM Users Manual.
f.
Verify the DMM error at 35.0 V, -35.0 V, 350.0 V (use
). Make sure the
errors are in specification limits. When
makes the output to go over 33 V,
the Calibrator goes into standby. When this occurs, push
to operate.
g.
Verify the DMM error at 1000 V to make sure it is in specification limits.
h.
Set the output of the Calibrator to 350 mV and push
. Make sure that the
errors are in specification limits.
4.
Verify the ac voltage function:
a.
Push
on the Calibrator and set the DMM function switch to
.
b.
Set the output of the Calibrator to 350 mV at 60 Hz and push
. Make sure
the errors are in specification limits.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...