5502A
Operators Manual
10-28
Verification Tables
At the Fluke Calibration factory, the SC300 Option is verified to its specifications at the
subsequent test points. The verification test points are given here as a guide when
verification is necessary.
Table 10-1. Voltage Function Verification: AC Voltage into a 1 M
Ω
Load
Nominal Value (p-p)
Frequency
Measured Value (p-p)
Deviation
(mV)
1-Year
Spec.
5.0 mV
10 Hz
0.11 mV
5.0 mV
100 Hz
0.11 mV
5.0 mV
1 kHz
0.11 mV
5.0 mV
5 kHz
0.11 mV
5.0 mV
10 kHz
0.11 mV
10.0 mV
10 kHz
0.12 mV
20.0 mV
100 Hz
0.15 mV
20.0 mV
1 kHz
0.15 mV
20.0 mV
10 kHz
0.15 mV
50.0 mV
10 kHz
0.23 mV
89.0 mV
10 Hz
0.32 mV
89.0 mV
10 kHz
0.32 mV
100.0 mV
10 kHz
0.35 mV
200.0 mV
100 Hz
0.60 mV
200.0 mV
1 kHz
0.60 mV
200.0 mV
10 kHz
0.60 mV
500.0 mV
10 kHz
1.35 mV
890.0 mV
10 Hz
2.32 mV
890.0 mV
10 kHz
2.32 mV
1.0 V
100 Hz
2.60 mV
1.0 V
1 kHz
2.60 mV
1.0 V
10 kHz
2.60 mV
2.0 V
10 kHz
5.10 mV
5.0 V
10 Hz
12.60 mV
5.0 V
10 kHz
12.60 mV
10.0 V
10 kHz
25.10 mV
20.0 V
10 kHz
50.10 mV
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...