Front-Panel Operation
Set Output
4
4-31
(upper reading) and AUX terminals (lower reading) (see below).
nn089f.eps
14.
Push
. The Calibrator erases your entry from the Control Display and copies
it into the Output Display (below is typical).
nn090f.eps
15.
Push
to activate the Calibrator output.
A softkey named “LO”s shows on the Control Display (see below).
nn091f.eps
•
“LO”s
(Low Potential Output Terminals) The front panel NORMAL LO and AUX
LO terminals must be tied together either at the UUT or at the Calibrator. When the
front panel NORMAL LO and AUX LO terminals are tied at the UUT, select “open”
with the “LO”s softkey. If the NORMAL LO and AUX LO terminals are not tied at
the UUT, select “tied” with the “LO”s softkey. The default is tied.
Set a Dual AC Voltage Output
Note
Tie the terminals NORMAL LO and AUX LO together at the UUT or at the
Calibrator, with the “LO”s softkey selection “tied.”
The Calibrator makes a dual ac voltage output by sourcing one ac voltage on the
NORMAL outputs and a second on the AUX terminals.
Do the subsequent procedure to set a dual ac voltage output. If you make an entry error,
push
one or more times to clear the display, then enter the value again.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...