5502A
Operators Manual
4-40
4.
Push the numeric keys and decimal point key to record the necessary temperature
output (maximum six numeric keys). For example, 123.456.
5.
For an output in
°
C, push
. For
°
F, push
and then
.
6.
The Control Display shows the amplitude of your temperature output. For example,
123.456
°
C (see below).
nn102f.eps
7.
Push
. The Calibrator erases your entry from the Control Display and copies
it into the Output Display (below is typical).
nn103f.eps
8.
Push
to activate the Calibrator output.
Four softkey names show on the Control Display. Push the OUTPUT softkey to toggle
the rtd selection, this shows the rtd setup menu and four softkey positions.
Note
The temperature you recorded above will be changed to 0
°
C (32
°
F) if you
change between tc (thermocouple) and rtd (resistance temperature
detector), or change the type of rtd. If this occurs, select OUTPUT rtd, the
necessary rtd TYPE, and then record the temperature again. Follow steps 4
to 8.
nn105f.eps
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...