Remote Operation
Set up the RS-232 Host Port for Remote Control
5
5-13
4.
Open Terminal from the Accessory group of Program Manager (below). If a terminal
configuration file already exists, for example, host.trm, select the necessary file with
the Open command from the File menu and go to Step 7. If not, go to Step 5.
nn308f.bmp
5.
Select the Communications command from the Setting menu. Record the RS-232
parameters that match those selected at the Calibrator for the Host port. If you use the
Calibrator factory defaults, you will see the Communications dialog box for COM1
as shown below. Select COM as necessary. Click OK.
nn309f.bmp
6.
Verify the Calibrator is powered and in the reset condition. (If in doubt, push
on
the Calibrator front panel.)
7.
On the Terminal screen, type the command REMOTE and push <Enter>. See that the
Calibrator Control Display changes to REMOTE CONTROL (see below).
nn325f.eps