Remote Commands
Commands
6
6-21
ISCR0?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
Error!
Bookmark not defined.
(Instrument Status 1 to 0 Change Register query) Shows and clears the contents of the
Instrument Status 1 to 0 Change Register.
Response:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCRO?
shows
6272
Shows decimal
6272
(binary
0001010001000000
) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
ISCR1?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
Error!
Bookmark not defined.
(Instrument Status 0 to 1 Change Register query) Shows and clears the contents of the
Instrument Status 0 to 1 Change Register.
Response:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISCR1?
shows
6272
Shows decimal
6272
(binary
0001010001000000
) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
ISR?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
Error!
Bookmark not defined.
(Instrument Status Register query) Shows contents of the Instrument Status Register.
Response:
<value> (decimal equivalent of the 16 bits, 0 to 32767)
Example:
ISR
? shows
6272
Shows decimal
6272
(binary
0001010001000000
) if bits 12 (SETTLED), 10
(REMOTE), and 6 (HIVOLT) are set to 1.
LCOMP
Overlapped
Coupled
IEEE-488
RS-232
Sequential
x
x
x
x
x
Error!
Bookmark not defined.
(Inductive compensation command) Activates or deactivates inductive load
compensation for ac current output. For current output, compensation is permitted when
the frequency is less than 440 Hz and the amplitude is less than 0.33 A. Compensation is
also permitted when the frequency is less than 1 kHz and the amplitude is larger than or
equal to 0.33 A.
Parameters:
OFF
(turns off the inductive load compensation circuitry)
ON
(turns on the inductive load compensation circuitry
Example:
LCOMP
ON
LCOMP?
Sequential
Overlapped
Coupled
IEEE-488
RS-232
x
x
x
x
x
Error!
Bookmark not defined.
(Inductive compensation query) Shows if inductive load compensation for ac current
output is active.
Responses: (character)
OFF
(Inductive load compensation circuitry is off)
(character)
ON
(Inductive load compensation circuitry is on)
Example:
LCOMP?
shows
ON
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...