Remote Operation
Set up the IEEE-488 Port for Remote Control
5
5-7
IEEE-488 Port Setup Procedure
Complete the subsequent procedure to set up the Calibrator for remote operations with
the IEEE-488 remote control port. This procedure selects GPIB as the interface and to
select the GPIB address for the interface.
1.
Turn the Calibrator power on. You can operate the Calibrator during warmup, but
specifications are not guaranteed until the warmup is complete.
2.
Push
on the Calibrator front panel.
3.
Navigate the softkey selections shown below. Verify the HOST port selection is
GPIB. Select the necessary GPIB port address (0 to 30) with the UP/DOWN
softkeys. The factory default is 4.
Select
nn120f.eps
4.
Push
(not
) several times until the message STORE
CHANGES/DISCARD CHANGES appears or, if there were no changes, the reset
display. If you select STORE CHANGES, the GPIB and host port information are
kept in the instrument non-volatile memory.