5502A
Operators Manual
4-46
Set Harmonics
When the Calibrator outputs dual ac voltages or ac power (sine waves only), the
Calibrator sources two signals with adjustable harmonic difference, with a maximum
harmonic frequency output of 10 kHz. For example, a 120 V, 60 Hz signal can be set on
the front-panel NORMAL terminals, and a 1 V, 300 Hz (5th harmonic) output on the
AUX terminals. The fundamental can be configured on the NORMAL or the AUX
terminals, with the harmonic output on the opposite terminals. Note that the maximum
AUX output is 3.3 V, while the maximum NORMAL output is 1000 V. Unless the
fundamental and harmonic frequencies are permitted for the given amplitude, the output
is not permitted.
Complete the subsequent procedure to record a harmonic output. For this procedure, you
will have already sourced a dual ac voltage or ac power output.
1.
Push the softkey WAVE MENUS to open the waveform menu.
2.
Push the softkey HARMONIC MENUS to open the harmonic submenu (below is
typical).
nn108f.eps
3.
Push the softkey FUNDMTL to select the Calibrator front-panel terminals for the
fundamental output (NORMAL or AUX). The harmonic shows on the AUX
terminals.
Summary of Contents for 5502A
Page 4: ......
Page 12: ...5502A Operators Manual xii ...
Page 46: ...5502A Operators Manual 2 2 ...
Page 52: ...5502A Operators Manual 2 8 ...
Page 54: ...5502A Operators Manual 3 2 ...
Page 142: ...5502A Operators Manual 4 68 ...
Page 194: ...5502A Operators Manual 6 2 ...
Page 240: ...5502A Operators Manual 7 2 ...
Page 260: ...5502A Operators Manual 7 22 ...
Page 262: ...5502A Operators Manual 8 2 ...
Page 365: ...B 1 Appendix B ASCII and IEEE 488 Bus Codes ...
Page 366: ...5502A Operators Manual B 2 ...
Page 368: ...5502A Operators Manual B 4 ...
Page 378: ...5502A Operators Manual D 6 ...